Device recall evidence
Joined enrichment rows that support device recall linkage and root-cause wording.
Indexed record set
39,516 evidence rows
-
- Evidence
- device-recall:cfres:195370
- Product
- Dimension Vista LOCI CA 15-3 Flex reagent cartridge. In Vitro Diagnostic for the quantitative measurement of CA 15-3. Siemens Material Number (SMN)/Ref Catalog Number : 10464344/K6456
- Join
- Official recall number ·
Z-1648-2022 - Root cause
- Under Investigation by firm
-
- Evidence
- device-recall:cfres:195615
- Product
- BEQ-HLS 7050 USA; HLS Set Advanced 7.0 Part No. 70106.9078. A preconnected set used for extracorporeal respiratory and/or cardiovascular support, as part of the CARDIOHELP System.
- Join
- Official recall number ·
Z-1767-2022 - Root cause
- Package design/selection
-
- Evidence
- device-recall:cfres:195663
- Product
- AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US
- Join
- Official recall number ·
Z-1769-2022 - Root cause
- Under Investigation by firm
-
- Evidence
- device-recall:cfres:195989
- Product
- Full Kit Pulset 3cc Syringe 23g x 1 Safety Tip 25u BH, REF Number 3373-91
- Join
- Official recall number ·
Z-1779-2022 - Root cause
- Process control
-
- Evidence
- device-recall:cfres:194779
- Product
- MEDLINE MICRO-KILL GERMICIDAL BLEACH WIPES, Model numbers MSC351420AN, MSC351420ANH, MSC351420ANZ, MSC351430A, MSC351430AZ
- Join
- Official recall number ·
Z-1493-2022 - Root cause
- Under Investigation by firm
-
- Evidence
- device-recall:cfres:194238
- Product
- Legacy Base for Dual Bed Configurations
- Join
- Official recall number ·
Z-1500-2022 - Root cause
- Component design/selection
-
- Evidence
- device-recall:cfres:195555
- Product
- ZAP-X Radiosurgery System
- Join
- Official recall number ·
Z-1742-2022 - Root cause
- Software design
-
- Evidence
- device-recall:cfres:195730
- Product
- PTS Panels CHOL+GLU Test Strips (REF 1765)
- Join
- Official recall number ·
Z-1741-2022 - Root cause
- Labeling Change Control
-
- Evidence
- device-recall:cfres:195920
- Product
- FABCO ORS DENTAL ROLLS, 1-1/2" STERILE, COTTON 3/8" Diameter x 1-1/2" (9.5mm x 38.1mm), packaged 5 /Pouch, 50 Pouches/Box Reorder Number: 34911510
- Join
- Official recall number ·
Z-1747-2022 - Root cause
- Component change control
-
- Evidence
- device-recall:cfres:195381
- Product
- TRVL KIT TOTHBRSH CAP PASTE IN POUCH 1CT, SKU 900737
- Join
- Official recall number ·
Z-1535-2022 - Root cause
- Storage
-
- Evidence
- device-recall:cfres:195155
- Product
- Wolf-Pak Premium Dressing Kit with GuardVA Cs/30 Ref: DC5030LF
- Join
- Official recall number ·
Z-1549-2022 - Root cause
- Labeling design
-
- Evidence
- device-recall:cfres:195389
- Product
- GS SALINE SOLUTION 12FL OZ, SKU 902274
- Join
- Official recall number ·
Z-1543-2022 - Root cause
- Storage
-
- Evidence
- device-recall:cfres:195386
- Product
- OPTI FREE REPLENISH MULTI PURPOSE 2 OZ
- Join
- Official recall number ·
Z-1540-2022 - Root cause
- Storage
-
- Evidence
- device-recall:cfres:195378
- Product
- CURAD FIRST AID KIT, SKU 999037
- Join
- Official recall number ·
Z-1532-2022 - Root cause
- Storage
-
- Evidence
- device-recall:cfres:194157
- Product
- Ethicon SURGIFLO Hemostatic Matrix Kit with Thrombin (6/CS) Ref: 2994 Mfg: FERROSAN MEDICAL DEVICES A/S Indicated in surgical procedures (other than ophthalmic) as an adjunct to hemostasis when control of bleeding by ligature or other conventional methods is ineffective or impractical.
- Join
- Official recall number ·
Z-1584-2022 - Root cause
- Environmental control
-
- Evidence
- device-recall:cfres:194442
- Product
- Drugs of Abuse Tests: a. Wondfo Tdip Drugs of Abuse Test Tdip Single Drug Fentanyl (FEN) Urine Sample 25 Tests. Model Number: AWFTY-114. b. Alere Drugs of Abuse Test E-Z Split Key Cup 12-Drug Panel AMP, BAR, BUP, BZO, COC, mAMP/MET, MDMA, MOP, MTD, OXY, PPX, THC Urine Sample 25 Tests Model Number: DOA-1127-041-19. c. Instant Technologies, Inc. Drugs of Abuse Test iCup A.D. 5-Drug Panel with Adulterants AMP, COC, mAMP/MET, OPI, THC (OX, pH, SG) Urine Sample 25 Tests Model Number: I-DUA-157-013. d. Instant Technologies, Inc. Drugs of Abuse Test iCup A.D. 6-Drug Panel with Adulterants AMP, BZO, COC, mAMP/MET, OPI, THC, (OX, pH, SG) Urine Sample Tests Model Number: I-DUA-167-022. e. Alere Drugs of Abuse Test iCup Dx Pro 2 10-Drug Panel with Adulterants AMP, BAR, BUP, BZO, COC, mAMP/MET, OPI300, OXY, MTD, THC (CR, OX, SG) Urine Sample 25 Tests Model Number: I-DXP-2107-01. f. Alere Drugs of Abuse Test iScreen 5-Drug Panel AMP, COC, mAMP/MET, OPI, THC Urine Sample 25 Tests Model Number: IS5M. g. Instant Technologies, Inc. Drugs of Abuse Test T-Cube 7-Drug Panel AMP50, COC20, mAMP/MET50, OPI40, OXY20, PCP10, THC40 Saliva Sample 25 Tests Model Number: TCUBE-7PCPB. h. Wondfo Drugs of Abuse Test with Adulterants Tcup 14-Drug Panel AMP1000, BAR300, BZO300, BUP10, COC300, mAMP/MET1000, MDMA500, MTD300, MOP(OPI300), OXY100, PCP25, PPX300, TCA1000, THC50 (CR, pH, SG) Urine Sample 25 Tests Model Number: TDOA-1145A3. i. Instant Technologies, Inc. Drugs of Abuse Test UScreen 12-Drug Panel with Adulterants AMP, BAR, BZO, COC, mAMP/MET, MDMA, MOP, MTD, OXY, PCP, TCA, THC, (CR, pH, SG) Urine Sample 25 Tests Model Number: USSCUPA-12CLIA. j. Instant Technologies, Inc. Drugs of Abuse Test UScreen 12-Drug Panel with Adulterants AMP, BAR, BUP, BZO, COC, mAMP/MET, MDMA, MOP, MTD, OXY, PCP, THC (CR, pH, SG) Urine Sample 25 Tests Model Number: USSCupA-12BUP300. k. Alere Drugs of Abuse Test E-Z Split Key Cup 6-Drug Panel AMP, BZO, COC, mAMP/MET, OPI, THC Urine Sample 25 Tests Model Number: DOA-2167-019. l. Alere Drugs of Abuse Test E-Z Split Key Cup A.D. 5-Drug Panel with Adulterants AMP, COC, mAMP/MET, OPI, THC (CR, GL, NI, OX, pH, SG) Urine Sample 25 Tests Model Number: DUD-157-012-019. m. WONDFO USA CO., LTD. Wondfo Drug Test Dip Card Drugs of Abuse Test Tdip Single Drug Buprenorphine (BUP) Urine Sample 25 Tests Model Number: WDBU-114.
- Join
- Official recall number ·
Z-1589-2022 - Root cause
- Environmental control
-
- Evidence
- device-recall:cfres:195311
- Product
- Servo-air Ventilator System-The Servo-air Ventilator System is: Intended for respiratory support, monitoring and treatment of pediatric and adult patients. Model Number: 6882000 (equipped with 6881999 mobile cart)
- Join
- Official recall number ·
Z-1568-2022 - Root cause
- Device Design
-
- Evidence
- device-recall:cfres:195695
- Product
- Artis Q floor- Model No. 10848280 Artis Q ceiling- Model No. 10848281 Artis Q biplane- Model No. 10848282 Artis Q zeego- Model No. 10848283 Artis Q.zen floor- Model No. 10848353 Artis Q.zen ceiling- Model No. 10848354 Artis Q.zen biplane- Model No. 10848355
- Join
- Official recall number ·
Z-1645-2022 - Root cause
- Device Design
-
- Evidence
- device-recall:cfres:195164
- Product
- NeuRx Diaphragm Pacing System (NeuRx DPS) External Pulse Generator (EPG), Part No. 23-0021.
- Join
- Official recall number ·
Z-1561-2022 - Root cause
- Nonconforming Material/Component
-
- Evidence
- device-recall:cfres:194612
- Product
- MEDLINE INDUSTRIES Medline Plug, Catheter Medline Plastic, with Cover Model Number: DYND12200
- Join
- Official recall number ·
Z-1610-2022 - Root cause
- Environmental control
-
- Evidence
- device-recall:cfres:195403
- Product
- Nobel Biocare N1 TiUltra TCC NP 3.5x13mm
- Join
- Official recall number ·
Z-1565-2022 - Root cause
- Process change control
-
- Evidence
- device-recall:cfres:195034
- Product
- VITEK 2 / MYLA VITEK 2 Software versions 8.01, 8.02 (veterinary use only), 9.01, 9.02, 9.03 MYLA Software versions V4.8 and V4.9
- Join
- Official recall number ·
Z-1640-2022 - Root cause
- Software design
-
- Evidence
- device-recall:cfres:194747
- Product
- a. McKesson DISINFECTANT, GLUTARALDEHYDE 28DAY (4GL/CS) MGM68, Model Number: 68-102800. b. METREX RESEARCH McKesson 14-Day Glutaraldehyde Solution DISINFECTANT, GLUTARALDEHYDE 14DAY (4GL/CS) MGM68 Model Number: 68-101400. c. METREX RESEARCH McKesson SOLUTION, GLUT 2.5% 28DAY QT (16/CS) PSS744 Catalog # 341. d. METREX RESEARCH LLC MetriCide DISINFECTANT, METRICIDE 14DAYGLUTARALDEHYDE 1GL (4/CS) Catalog # 10-1400. e. METREX RESEARCH LLC MetriCide28 DISINFECTANT, METRICIDE GLUT LNG LIFE GL (4GL/CS) Catalog # 10-2800
- Join
- Official recall number ·
Z-1614-2022 - Root cause
- Environmental control
-
- Evidence
- device-recall:cfres:195532
- Product
- Wisp and Wisp Youth Nasal Mask
- Join
- Official recall number ·
Z-1783-2022 - Root cause
- Error in labeling
-
- Evidence
- device-recall:cfres:195752
- Product
- Cope Mandril Wire Guide (Nitinol), Nitinol Guidewire; Reference Part Number/GPN: PMG-18SP-60-COPE-NT-ST G09395 PMG-18SP-100-COPE-NT G08687 PMG-18SP-60-COPE-NT G08427
- Join
- Official recall number ·
Z-1829-2022 - Root cause
- Under Investigation by firm