Skip to content
Recall Observatory FDA recall evidence

Device product

AquaC UNO H [120V, 60Hz, 1000VA, US-Version - a reverse osmosis unit intended for use with hemodialysis systems. Fresenius Part Number: 24-10VA-1 Vivonic Part Number: G02000970-US

Z-1769-2022

June 21, 2022

Class II

Product summary

Firm
Fresenius Medical Care Holdings, Inc.
Event
Event 90757
Status
Ongoing
Classification
Class II
Quantity
24 units
Official record key
device-enforcement:Z-1769-2022

Official wording

Reason: Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

Code information: UDI: (01)4251285200234(11)210628(21)1CHS5379 (01)4251285200234(11)160927(21)6CHS1270 (01)4251285200234(11)161109(21)6CHS1394 (01)4251285200234(11)161109(21)6CHS1435 (01)4251285200234(11)170131(21)7CHS1569 (01)4251285200234(11)170201(21)7CHS1583 (01)4251285200234(11)170213(21)7CHS1619 (01)4251285200234(11)170314(21)7CHS1641 (01)4251285200234(11)170519(21)7CHS1790 (01)4251285200234(11)170807(21)7CHS1953 (01)4251285200234(11)180326(21)8CHS2197 (01)4251285200234(11)181010(21)8CHS2608 (01)4251285200234(11)181102(21)8CHS2633 (01)4251285200234(11)181107(21)8CHS2644 Serial Numbers: 1CHS5379 6CHS1192 6CHS1194 6CHS1195 6CHS1196 6CHS1197 6CHS1198 6CHS1199 6CHS1200 6CHS1201 6CHS1233 6CHS1270 6CHS1394 6CHS1435 7CHS1569 7CHS1583 7CHS1619 7CHS1641 7CHS1790 7CHS1953 8CHS2197 8CHS2608 8CHS2633 8CHS2644

Distribution pattern: US Nationwide distribution in the states of AZ, CA, DC, FL, GA, IN, MA, NV, NY, PA, TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Internal ethemet cables are missing one or both ferrite cores therefore the electromagnetic compatibility of the device is not ensured

Field note

Send feedback

We'll only use this to respond to your feedback.