Skip to content
Recall Observatory FDA recall evidence

Device product

AngioDynamics VenaCure EVLT NeverTouch Procedure Kit, 45CM NEVERTOUCH FRS .018 PROCEDURE KIT PG, Catalog Number 11403012

Z-2444-2019

June 27, 2019

Class II

Product summary

Firm
Angiodynamics, Inc.
Event
Event 83334
Status
Terminated
Classification
Class II
Quantity
109
Official record key
device-enforcement:Z-2444-2019

Official wording

Reason: The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

Code information: Lot 5456490; UPN H787114030120

Distribution pattern: The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.

Field note

Send feedback

We'll only use this to respond to your feedback.