openFDA Device Enforcement
Report-date coverage
June 20, 2012–July 08, 2026
Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
openFDA Device Recall
event_date_initiated coverage
June 01, 1997–June 30, 2026
Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.
Product dates, classifications, firms, and source wording remain attached to their individual rows.
Opening this dossier never hides products that did not match a prior timeline filter.
The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.
These labels are deterministic app interpretations, not FDA categories.
Official device-enrichment evidence · Unknown
An exact joined enrichment record exists, but none supplies supported root-cause wording. This is not an FDA finding of an unknown cause.
The affected product is properly labelled with an expiration date of March 31, 2022; but the RFID programmed expiration date is April-2014 which results in an error message when connected to the laser source and prevents the product from being used.
Code information
Lot 5456490; UPN H787114030120
Distribution pattern
The products were distributed to the following US states: DC, IL, IN, MI, NJ, TN, TX, and WI.