Skip to content
Recall Observatory FDA recall evidence

Device product

Identity Imprint PS Tibial Tray Size 4: Lot 540287

Z-1591-2024

March 11, 2024

Class II

Product summary

Firm
Conformis, Inc.
Event
Event 94288
Status
Ongoing
Classification
Class II
Quantity
3 units
Official record key
device-enforcement:Z-1591-2024

Official wording

Reason: Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.

Code information: Catalog Number: TPS-301-T04S; UDI#: M572TPS301T04S011, Lot number 540287

Distribution pattern: US Nationwide distribution in the states of FL, KS and NH.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.