Skip to content
Recall Observatory FDA recall evidence

Exact FDA event identifier

Event 94288

1 recalled-product record grouped only because every row carries this exact official event ID.

Open FDA event record

Event summary

Timeline bucket
March 11, 2024
Product types
Device
Classifications
Class II
Statuses
Ongoing
Recalling firm wording
Conformis, Inc.

Dossier provenance

Source snapshots represented here

  • openFDA Device Enforcement Report-date coverage June 20, 2012–July 08, 2026 Known gap: openFDA describes 2004-present coverage, but the current export has no report-date records in this interval.
  • openFDA Device Recall event_date_initiated coverage June 01, 1997–June 30, 2026

Enforcement coverage uses report dates; product initiation dates can precede those bounds. Device-enrichment coverage uses FDA event_date_initiated.

How coverage and grouping work

Complete imported group

Every recalled product in this event

1 official enforcement rows

Product dates, classifications, firms, and source wording remain attached to their individual rows. Opening this dossier never hides products that did not match a prior timeline filter.

device · product 1 of 1

Identity Imprint PS Tibial Tray Size 4: Lot 540287

Z-1591-2024
Recall number
Z-1591-2024
Initiated
March 11, 2024
Classification
Class II
Status
Ongoing
Recalling firm
Conformis, Inc.
Quantity
3 units

App-derived interpretation

Unknown reason.no_named_rule · v1.0.0
Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.

These labels are deterministic app interpretations, not FDA categories.

Official device-enrichment evidence · Sourced

Process design

Compare all device evidence states

Inspect official wording and provenance

Reason for recall

Size 5 Tibial Tray from Lot 508287 was packaged in a Size 4 box from Lot 540287. It was confirmed that Lot 540287 and Lot 508287 were swapped between the pouching and final pack.

Code information

Catalog Number: TPS-301-T04S; UDI#: M572TPS301T04S011, Lot number 540287

Distribution pattern

US Nationwide distribution in the states of FL, KS and NH.

Field note

Send feedback

We'll only use this to respond to your feedback.