Skip to content
Recall Observatory FDA recall evidence

Device product

Temperature Sensing 100% Silicone Foley Catheter, Part Number 102202101863DO - Product Usage: is a 3 lumen silicone catheter with a sensor to measure temperature while draining the urinary bladder.

Z-1786-2020

March 17, 2020

Class II

Product summary

Firm
Degania Medical Devices Pvt. Ltd.
Event
Event 85327
Status
Terminated
Classification
Class II
Quantity
2698
Official record key
device-enforcement:Z-1786-2020

Official wording

Reason: Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.

Code information: Lot Numbers: V19012401 V19012403

Distribution pattern: Worldwide distribution - US Nationwide distribution including state of IL and the country of Mexico.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Inadequate incoming inspection practices were utilized on thermistor sensors used in the production of the specified lots of temperature sensing catheters.

Field note

Send feedback

We'll only use this to respond to your feedback.