Device product

DxA 5000 (DxA Automation System, DxA 5000 fit), "For in Vitro Diagnostic Use", Reference Numbers: B87352 with B57634, UDI: 15099590707118

Z-2602-2021

July 21, 2021

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 88519
Status
Terminated
Classification
Class II
Quantity
123 systems
Official record key
device-enforcement:Z-2602-2021

Official wording

Reason: There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).

Code information: Affected US Serial Numbers: T23000077, T23000075, T23000076, T23000074, T23000069, T23000070, T23000066, T23000099, T23000096, T23000098, T23000100, T23000095, T23000097, T23000101, T23000092, T23000091, T23000090, T23000142, T23000045, T23000040, T23000039, T23000038, T23000152, T23000154, T23000155, T23000153, T23000106, T23000108, T23000107, T23000109, T23000128, T23000124, T23000125, T23000126, T23000115, T23000116, T23000113, T23000112, T23000117, T23000114, T23000132, T23000133, T23000131, T23000134, T23000064, T23000065, T23000063, T23000062, T23000163, T23000164, T23000162, T23000173, T23000169, T23000161, T23000138, T23000136, T23000137, T23000094, T23000093, T23000135, T23000141, T23000139, T23000140, T23000166, T23000168, T23000167, T23000082, T23000081, T23000080, T23000085, T23000084, T23000083, T23000143, T23000146, T23000145, T23000144

Distribution pattern: US Nationwide distribution in the states of AL, CA, IL, In, KY, MN, MI, OH, PA, TX, WV.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    There is a potential that sample carriers (w or w/o samples) within automated sampling system may transport samples to the outer lane which does not contain a physical barrier to restrain samples. Samples are dropped when they reach the end of the outer lane which could result in Delay in providing patient with appropriate medical treatment. New sample needs to be drawn. Substantial discharge of biohazards material - Infection with a human transmissible blood-borne or animal pathogen which could result in Infection with a human transmissible blood-borne pathogen (e.g. infectious HBV, HCV, HIV).