Skip to content
Recall Observatory FDA recall evidence

Device product

Albumin Gen.2, Catalog 05166861190

Z-0104-2021

August 27, 2020

Class III

Product summary

Firm
Roche Diagnostics Operations, Inc.
Event
Event 86406
Status
Terminated
Classification
Class III
Quantity
11963
Official record key
device-enforcement:Z-0104-2021

Official wording

Reason: Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Code information: lots 43031001, 43718901

Distribution pattern: Domestic distribution nationwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Roche has confirmed customer complaints of low quality (QC) recovery and invalid calibrations for the identified products. A discoloration (yellow color) was observed and is the result of cross contamination between the R1 and R3 reagents.

Field note

Send feedback

We'll only use this to respond to your feedback.