Skip to content
Recall Observatory FDA recall evidence

Device product

Abbott NT2000iX Radiofrequency (RF) generator, REF RFG-NT-2000, lesioning generator

Z-0209-2024

September 28, 2023

Class II

Product summary

Firm
Abbott
Event
Event 93152
Status
Ongoing
Classification
Class II
Quantity
34 devices
Official record key
device-enforcement:Z-0209-2024

Official wording

Reason: Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

Code information: UDI/DI 05415067022417, serial numbers: 16930509, 16983330, 17583407, 17634837, 17783928, 17783930, 17905961, 17928413, 17954521, 17991999, 18047384, 18493711, 18522937, 18675068, 18690920, 18709734, 18742997, 2475-12, 2504-12, 2547-13, 2885-14, 3003-14, 3278-15, 3450-15, 3550-15, 17942357, 18003310, 18102335, 18132017, 18132092, 18657521, 18666647, 19182119, 19197174

Distribution pattern: US, Taiwan, and Colombia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Certain NT2000iX RF generators were serviced using a tool that was subsequently found to be out of tolerance and therefore may not have provided accurate results for potential capacitance failure.

Field note

Send feedback

We'll only use this to respond to your feedback.