Skip to content
Recall Observatory FDA recall evidence

Device product

Omnicell IVX Station-For use in the automated preparation of pharmaceutical admixtures and syringes Model Number: 298921373

Z-1617-2023

March 27, 2023

Class II

Product summary

Firm
OmniCell, Inc.
Event
Event 92136
Status
Ongoing
Classification
Class II
Quantity
6 units
Official record key
device-enforcement:Z-1617-2023

Official wording

Reason: IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling

Code information: UDI-DI: 00860006285005 Serial Numbers: IVXS000014, IVXS000015, IVXS000016, IVXS000017, IVXS000018. IVXS000019

Distribution pattern: AL, NY

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    IVX Station does not account for inherent scale inaccuracy or variability, could result in a preparation being passed that is outside of the dose accuracy stated in the product labeling

Field note

Send feedback

We'll only use this to respond to your feedback.