Device product

EndoVive Replacement Button Kit w/ ENFIT 28F 2.8 CM, Outer Box UPN M00509500 or M00509510.

Z-1153-2024

December 19, 2023

Class II

Product summary

Firm
Boston Scientific Corporation
Event
Event 93794
Status
Ongoing
Classification
Class II
Quantity
0 US; 74 OUS
Official record key
device-enforcement:Z-1153-2024

Official wording

Reason: Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.

Code information: UPN M00509500, GTIN (UDI-DI) 08714729902713, Lot 31073152; UPN M00509510, GTIN (UDI-DI) 08714729902720, Lots 31221751, 31761882, 32409573.

Distribution pattern: International distribution to the country of Japan only.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Affected batches of the kit contain Stoma Measuring Devices sized to measure for an 18/24F replacement button, but are labeled as being for a 28F. Use of defective product may lead to undersizing the button, which may cause embedding with erosion into the gastric wall, tissue necrosis, infection, sepsis, and associate sequelae.