Skip to content
Recall Observatory FDA recall evidence

Device product

Arrow-Trerotola Over-The-Wire PTD Kit Percutaneous Thrombolytic Device : 7FR Product Code: PT-65709-HFWC The Arrow-Trerotola" Percutaneous Thrombolytic Device (PTDTM) Catheter, in conjunction with the Arrow Rotator Drive Unit, permits mechanical declotting of native arterio-venousfistulae and synthetic dialysis grafts

Z-0232-2022

September 20, 2021

Class I

Product summary

Firm
Arrow International Inc
Event
Event 88796
Status
Terminated
Classification
Class I
Quantity
1245 units
Official record key
device-enforcement:Z-0232-2022

Official wording

Reason: 7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.

Code information: Lot Number: 13F19K0369 13F20A0323 13F20B0139 13F20C0594 13F20F0083 13F20F0230 13F20F0577 13F20H0756 13F20K0849 13F20L0283 13F20M0182 13F21A0498 13F21B0158 13F21C0747 13F21E0555

Distribution pattern: Worldwide distribution: US (Nationwide) and OUS (Foreign) to countries of: BE CA CL CZ DE ES FR IL IT KR PH SA TR TW UK

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    7FR Percutaneous Thrombolytic Device (PTD) Inner lumen (orange) detached from the device basket result in embolization and retrieval with an additional intervention and consequent delay of therapy.

Field note

Send feedback

We'll only use this to respond to your feedback.