Skip to content
Recall Observatory FDA recall evidence

Device product

Omnipod 5 App (on compatible Android smartphones), Software Versions 1.1 - 1.2.3 (current). US: PT-000599- Omnipod 5 App Software Package (provided through Google Play), EU: PT-001323- Omnipod 5 App Software Package (provided on locked-down Controllers (PT-001194 and PT-001193). Used to control an insulin pump.

Z-0591-2024

November 30, 2023

Class I

Product summary

Firm
Insulet Corporation
Event
Event 93511
Status
Ongoing
Classification
Class I
Quantity
27000 users
Official record key
device-enforcement:Z-0591-2024

Official wording

Reason: The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.

Code information: UDI/Model/NDC Number: " Omnipod 5 App Software UDI: 1038508112030 " EU locked-down Controller UDI: 10385083000190 and 10385083000206

Distribution pattern: Worldwide distribution - US Nationwide and the countries of Germany and UK.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bolus calculator is not recording the decimal point if it is the first value entered when changing a bolus dose. This may lead to an over-delivery of insulin to the user if the user does not recognize the error on the bolus calculator screen or the confirmation screen prior to starting the bolus.

Field note

Send feedback

We'll only use this to respond to your feedback.