Skip to content
Recall Observatory FDA recall evidence

Device product

Torq-Flex Wire Guide Australian Modification, Guidewire Reference Part Number/GPN STF-18-40-AUST G07304

Z-1843-2022

August 16, 2022

Class II

Product summary

Firm
Cook Incorporated
Event
Event 90793
Status
Ongoing
Classification
Class II
Quantity
149 US
Official record key
device-enforcement:Z-1843-2022

Official wording

Reason: Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Code information: STF-18-40-AUST G07304 UDI-DI: 00827002073043 Lots/Expiration Date: 14788615 13-06-2027 14816230 28-06-2027 14836977 12-07-2027

Distribution pattern: Domestic distribution nationwide. Foreign distribution worldwide.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Cook Medical identified that devices from the affected device lots may have a complete breach of the chevron seal of the packaging. Therefore, the sterility of affected devices may be compromised.

Field note

Send feedback

We'll only use this to respond to your feedback.