Skip to content
Recall Observatory FDA recall evidence

Device product

Medtronic REF: MMT-7020D1 Guardian Sensor (3) for the following packaging: GS3 1PK OUS1/ GTIN: 00643169704084 GS3 1PK 9L OUS1/ GTIN: 00763000243333 GS3 1PK 9L OUS1/ GTIN: 00763000318857 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

Z-1612-2021

February 26, 2021

Class II

Product summary

Firm
Medtronic Inc.
Event
Event 87476
Status
Ongoing
Classification
Class II
Quantity
111,678 units
Official record key
device-enforcement:Z-1612-2021

Official wording

Reason: Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Code information: All Lot numbers within active shelf life

Distribution pattern: Worldwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Use of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.

Field note

Send feedback

We'll only use this to respond to your feedback.