Skip to content
Recall Observatory FDA recall evidence

Device product

BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Z-1018-2022

March 09, 2022

Class II

Product summary

Firm
Becton Dickinson & Co.
Event
Event 89896
Status
Terminated
Classification
Class II
Quantity
15 systems
Official record key
device-enforcement:Z-1018-2022

Official wording

Reason: When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

Code information: All systems running BD Synapsys version 4.10; UDI: 00382904441500; Serial Numbers: SL00530 SL00515 SL00527 SL00502 SL00540 SL00564 SL00549 SL00509 SL00003 SL00004 SL00010 SL00001 SL00005 SL00002 SL000015 SL01010 SL01022

Distribution pattern: Worldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

Field note

Send feedback

We'll only use this to respond to your feedback.