Skip to content
Recall Observatory FDA recall evidence

Device product

Arcos Modular Revision Hip System Standard Cone Prox Body, Size C, 80 MM Item Number: 11-301343

Z-2274-2021

June 02, 2021

Class II

Product summary

Firm
Biomet, Inc.
Event
Event 88304
Status
Terminated
Classification
Class II
Quantity
N/A
Official record key
device-enforcement:Z-2274-2021

Official wording

Reason: Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention

Code information: Lot Numbers/UDI : 626750 (01)00880304474499(17)310415(10)626750

Distribution pattern: US Nationwide distribution.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Arcos Proximal Cone Bodies could potentially exhibit chatter in the inner taper fretting corrosion and device failure in vivo leading to surgical intervention