Device product
Cascadion SM Clinical Analyzer, REF 99990000 SN Type: 9910
Z-1202-2021
Product summary
- Event
- Event 87309
- Status
- Terminated
- Classification
- Class II
- Quantity
- 19 systems (1 system in U.S. and 10 system O.U.S.)
- Official record key
device-enforcement:Z-1202-2021
Official wording
Reason: Due to software defect, under certain assay parameters, false Vitamin D results may be reported. The system reports a false result by not quantitating the correct analyte peak in sample chromatogram. The resulting sample would show a Vitamin D-concentration that is an unusual situation for human serum and plasma samples and would indicate a severe Vitamin D deficiency.
Code information: Model Number 99990000 GTIN: 06438153004925 Serial Numbers: CA100077, CA000055, CA100073, CA100074, CA000051, CA100062, CA100065, CA100069, CA100070 CA100063 CA000042, CA100071 CA100060, CA100059, CA000058, CA000053, CA000056, CA100061, CA100066
Distribution pattern: U.S.: Texas O.U.S.: Belgium, France, Germany, Italy, Spain, Switzerland and United Kingdom
Derived failure modes
-
Device software or design
software defect