RANDOX CHOL-cholesterol (total) test system intended to measure cholesterol in plasma and serum REF: CH200

Z-0328-2022

November 01, 2021

Class II

Product summary

Firm
Randox Laboratories Ltd.
Event
Event 88999
Status
Terminated
Classification
Class II
Quantity
15 kits
Official record key
device-enforcement:Z-0328-2022

Official wording

Reason: When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range

Code information: Lot Numbers/Exp Date: 559344 28th June 2022; 572730 28th July 2023. GTIN: 05055273201130

Distribution pattern: US Nationwide distribution in the states of GA, OR, PA ,WA, West Virginia.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    When calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range