Skip to content
Recall Observatory FDA recall evidence

Device product

Bien-Air Surgery REF 1600407-001 SN **** FOOTCTRL OSSEODOC - Product Usage: intended to convert electrical energy to mechanically drive handpieces and microsaws used to cut and shape bones.

Z-0176-2021

September 11, 2020

Class II

Product summary

Firm
Bien-Air Surgery Sa
Event
Event 86419
Status
Ongoing
Classification
Class II
Quantity
Total US = 1120 foot pedals
Official record key
device-enforcement:Z-0176-2021

Official wording

Reason: The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.

Code information: Model 1600407 / Catalog number 1600407-001 Serial Numbers: 16C0017 16F0013 16I0009 16K0017 17F0039 17I0015 17L0013

Distribution pattern: US Nationwide distribution including in the states of CA, MI and MN.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The magnet located inside the foot pedal may come off and may result in unattended start of unstoppable motor.

Field note

Send feedback

We'll only use this to respond to your feedback.