Skip to content
Recall Observatory FDA recall evidence

Device product

Proteus 235 The Proton Therapy System - Proteus 235 (brand names: Proteus Plus and Proteus ONE) is a medical device designed to produce and deliver a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Z-2715-2020

July 17, 2020

Class II

Product summary

Firm
Ion Beam Applications S.A.
Event
Event 86076
Status
Terminated
Classification
Class II
Quantity
2 Units
Official record key
device-enforcement:Z-2715-2020

Official wording

Reason: IBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.

Code information: Serial Number: Manufacture date SAT.140 (US): manufactured May 2019 SBF.112 (EU): manufactured October 2019

Distribution pattern: US: VA OUS: Belgium

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    IBA became aware of an issue to resume aborted irradiations because the plan to complete the treatment cannot be loaded.

Field note

Send feedback

We'll only use this to respond to your feedback.