Skip to content
Recall Observatory FDA recall evidence

Device product

Neo-Vent Infant T-Piece Resuscitation Circuit;Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0415, Qty:10, Length: 60" (1.5cm), Rx Only.

Z-2434-2020

January 15, 2019

Class II

Product summary

Firm
Westmed, Inc.
Event
Event 85576
Status
Terminated
Classification
Class II
Quantity
625 cases
Official record key
device-enforcement:Z-2434-2020

Official wording

Reason: A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Code information: 101118T42, 101118T44, 102918T01, 080618T02, 080618T03 , 101118T43, 102918T02.

Distribution pattern: Worldwide distribution - US Nationwide distribution including in the states of CA, GA, IL,OH, MA, MI, MO, NC, OH, OR, TX, VA, and WI. The country of Spain.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Field note

Send feedback

We'll only use this to respond to your feedback.