Vista Dental Products CanalClean Endodontic Irrigation Kit, (1) 12mL - Chlor-XTRA, 6mL - SmearOFF + 92) + (2) 30ga Irrigation Tips + (1) Micro Aspirator with Tip, REF 504700

Z-0011-2020

June 27, 2019

Class II

Product summary

Firm
Inter-Med Llc
Event
Event 83435
Status
Terminated
Classification
Class II
Quantity
377 units
Official record key
device-enforcement:Z-0011-2020

Official wording

Reason: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Code information: No UDI Lot/Work Order Numbers: 2018-1192 2018-1473 2018-1742 2018-1999 2018-2386 2018-2741 2019-0047

Distribution pattern: US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.