Skip to content
Recall Observatory FDA recall evidence

Device product

Vista Dental Products Sodium Hypochlorite - 6%, QTY: 12, 3mL Pre-filled Syringes, REF 502350

Z-0014-2020

June 27, 2019

Class II

Product summary

Firm
Inter-Med Llc
Event
Event 83435
Status
Terminated
Classification
Class II
Quantity
768 units
Official record key
device-enforcement:Z-0014-2020

Official wording

Reason: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Code information: UDI (01)10818207020458 Lot/Work Order Numbers: 20190197 20190723 20190834 20190862 20190952

Distribution pattern: US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Field note

Send feedback

We'll only use this to respond to your feedback.