VITROS XT 7600 Integrated System, with V3.4 or 3.4.1 SW Product Code: 6844461 Product Usage: INTENDED USE: For in vitro diagnostic use only. The VITROS XT 7600 Integrated System is intended for use in the measurement of a variety of analytes of clinical interest

Z-1364-2019

March 20, 2019

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 82561
Status
Terminated
Classification
Class II
Quantity
85 units: US=19;OUS=66
Official record key
device-enforcement:Z-1364-2019

Official wording

Reason: Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported

Code information: Serial Number 76000108 thru current.

Distribution pattern: Worldwide Distribution - US Nationwide in the states of AZ, CA, IL, OH, FL, TX, MI Foreign: Australia, Belgium, Chile, India, Italy, Japan, South Korea, Spain, Canada,, Phillippines

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for sample fluid to be dispensed to an incorrect position on the MicroSlide, potentially leading to erroneous assay results being reported