Vista Dental Products Sodium Hypochlorite - 3%, 10 - 12cc Pre-filled Syringes, REF 502315

Z-0013-2020

June 27, 2019

Class II

Product summary

Firm
Inter-Med Llc
Event
Event 83435
Status
Terminated
Classification
Class II
Quantity
582 units
Official record key
device-enforcement:Z-0013-2020

Official wording

Reason: The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.

Code information: Lot/Work Order Numbers: 2018-1349 2018-1587 2018-1853 20182107, UDI (01)10818207020427 20182303, UDI (01)10818207020427 20182584, UDI (01)10818207020427 20182666, UDI (01)10818207020427 20190199, UDI (01)10818207020427 20190433, UDI (01)10818207020427

Distribution pattern: US, South Africa, Switzerland, Poland, Italy, Singapore, Australia

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The affected lots may be packaged in syringes which include an incorrect bung. The incorrect bung is made of a latex-free isoprene rubber material. After prolonged storage and exposure to contained solutions, the rubber bung may degrade due to chemical interaction with the bleach solution.