Skip to content
Recall Observatory FDA recall evidence

Device product

EV1000 Clinical Platform, Models: EV1000A, EV1000NI, EV1000CS Product Usage: EV1000A Monitor is used to optimize fluid management through the use of minimally invasive monitoring with the FloTrac sensor (or the Volume View sensor).

Z-1193-2019

March 21, 2019

Class I

Product summary

Firm
Edwards Lifesciences, LLC
Event
Event 82456
Status
Terminated
Classification
Class I
Quantity
22443
Official record key
device-enforcement:Z-1193-2019

Official wording

Reason: Potential for an electrical short circuit leading to possible fire hazard due to liquid ingress into AC power outlets.

Code information: All lots

Distribution pattern: Worldwide Distribution - US Nationwide : FL, MD, NY, TX, VA, CO, OK, MO, CA, DE, TN, NC, DC, PA, MA, OR, SC, OH, WI, KS, NV, MI, IL, KY, GA, NJ, WA, CT, IN, MN, WV, LA, MS, AL, SD, NE, IA, AZ, ID, HI, UT, ME,MT, NM, AK, NH, RI, AR, ND, WY, PR. Foreign (OUS): IT, FI, CO, KR, JP, PL, BE, GR, NL, IL, AT, DK,CA, GB, BR, SA, SE, DE, AU, FR, MX, IN, IC, ZA, ES, TR, NZ, MY, IE, SG, CZ, SI, PT, CH, RE, LK, CR, AR, HK, NO, PY, PA, MT, EG, NP, TN, RU, RO, NA,CL, BH, IR, PH, TH, KZ, DZ, BG, EE, KW, VN, OM, PE, EC, AE, MA, VE, BO, QA, DO, GT, BY, JO, LT, SK, HU, CY, MN, TW, CN

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential for an electrical short circuit leading to possible fire hazard due to liquid ingress into AC power outlets.

Field note

Send feedback

We'll only use this to respond to your feedback.