Device product
AVE 2 Birthing Bed
Z-0991-2019
Product summary
- Event
- Event 81927
- Status
- Terminated
- Classification
- Class II
- Quantity
- 253 devices
- Official record key
device-enforcement:Z-0991-2019
Official wording
Reason: LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.
Code information: Serial Numbers 4PPB0091 through 4PPB626
Distribution pattern: AL, OK, NJ, OR, WA, CA, MN, LA, AZ, IL, MO, CO, OH, KY, FL, KS, UT, TN, LA, DE, CT, TX
Derived failure modes
-
Unknown
LINET has recently become aware of a potential safety issue for the AVE 2 Birthing Bed caused by the power cable breaking at the bed connection or the orange insulation of the power cable may fail causing a possible spark or possible exposure to a live conductor.