Device product

Cobas connection module (CCM) - OUTPUT UNIT The module CCM enables a unidirectional connection of the pre-analytical systems cobas p 612/ p 512 to analysis instruments from Roche Diagnostics, Hitachi or Sysmex by means of the Aloka transportation system

Z-2223-2017

May 09, 2017

Class II

Product summary

Firm
Roche Diagnostics Corporation
Event
Event 77243
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-2223-2017

Official wording

Reason: a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.

Code information: Serial Number 204C5690 204H0157 204H0158

Distribution pattern: US distribution to WA only.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    a gap between the conveyor belt and the rack tray table (plastic part) may cause vibration of the 5 position racks during the transport. This issue can pose a potential risk to involved operators and can also lead to potential cross contamination of samples in the affected racks.