Skip to content
Recall Observatory FDA recall evidence

Device product

VOLISTA StandOP Surgical Light, ARD568811961 VLT600DF AIM STP; ARD568821961 VLT600DF AIM STP QL; ARD568811911 VLT600SF AIM STP; ARD568821911 VLT600SF AIM STP QL; ARD568821910 VLT600SF STP QL; ARD568811901 VLT600SF AIM STP Product Usage: The Maquet VOLISTA surgical lights are designed to illuminate the body of a patient during surgical operations, diagnostics and treatment. StandOP VOLISTA is a system of ceiling-mounted surgical lights suited for installation in surgical suites, examining rooms, doctors surgeries and out-patient consultations.

Z-2502-2018

June 14, 2018

Class II

Product summary

Firm
Maquet SAS
Event
Event 80452
Status
Terminated
Classification
Class II
Quantity
151
Official record key
device-enforcement:Z-2502-2018

Official wording

Reason: Certain VOLISTA StandOP Surgical Lights have been identified as having a potential light head detachment issue. Improper assembly of the bracket that connects the light head to the axle fork could result in the light head bracket breaking, causing the light head to detach from the arm and become suspended only from the cabling inside the arm. Use of an affected device may result in serious injury to patient and/or medical staff if the light head detaches and drops, impacting patient and/or medical staff during demonstration, service, or use.

Code information: All Serial Numbers manufactured from March 10, 2014 until November 17, 2016 6.

Distribution pattern: Worldwide Distribution - US Nationwide in the states of AR, AZ, CA, CO, CT, FL, GA, IA, ID, KS, KY, LA, MD, ME, MO, NE, NY, OH, OR, TN, TX, UT, and WA. Foreign distribution to Argentina, Australia, Austria, Bangladesh, Belgium, Bolivia, Brazil Cameroon, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, Germany, Ghana, Hong Kong, Hungary, India, Iran, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Namibia, Netherlands, Nigeria, Oman, Pakistan, Panama, Paraguay, Poland, Portugal, South Korea, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, United Kingdom, Tanzania, Yemen, and Zimbabwe.

Derived failure modes

  • Manufacturing or process control

    reason.manufacturing_process_control · v1.0.0

    Improper assembly

Field note