Skip to content
Recall Observatory FDA recall evidence

Device product

iChem VELOCITY Automated Urine Chemistry System, Catalog #700-7176-001 and iChem VELOCITY Automated Urine Chemistry System Computerless, Catalog #700-7177-001. The iChemVELOCITY automated urine chemistry system is an in-vitro diagnostic device used to automate the urine chemistry analysis profile using iChemVELOCITY Urine Chemistry Strips. The iChemVELOCITY can be used as a stand-alone system, as well as in a iQ200 Series system, a configuration given the proprietary name iRICELL as it is designed to be hardware and software compatible with iQ200 Series systems.

Z-1894-2018

March 29, 2018

Class II

Product summary

Firm
Beckman Coulter Inc.
Event
Event 79895
Status
Terminated
Classification
Class II
Quantity
6 devices
Official record key
device-enforcement:Z-1894-2018

Official wording

Reason: The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.

Code information: Catalog 700-7176-001 - Serial numbers V04494 and CV04015; Catalog 700-7177-001 - Serial numbers CV04012, CV04013, CV04014, and CV04017.

Distribution pattern: US Distribution to states of: AL, IA, GA, and FL.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The Strip Provider Modules (SPM) on some of the instruments had not been inspected during manufacture for correct seal placement and integrity.

Field note

Send feedback

We'll only use this to respond to your feedback.