Skip to content
Recall Observatory FDA recall evidence

Device product

VITROS Chemistry Products URIC Slides, REF/Product Code 194 3927 (300 slides per sales unit), Unique Device Identifier No. 10758750000340, Rx ONLY, IVD --- For in vitro diagnostic use only.

Z-0092-2017

August 11, 2016

Class II

Product summary

Firm
Ortho-Clinical Diagnostics
Event
Event 74950
Status
Terminated
Classification
Class II
Quantity
US: 18,749 sales units; Foreign: 94,928 sales units
Official record key
device-enforcement:Z-0092-2017

Official wording

Reason: Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.

Code information: Current/In-date GENs 26, 27, 28, 29, 31

Distribution pattern: Worldwide Distribution - US (Nationwide) Puerto Rico and Bermuda), Internationally to Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Poland, Singapore, Spain, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ortho Clinical Diagnostics confirmed that N-Acetylcysteine (NAC), a drug used for acetaminophen overdose, mucolytic therapy and is used as nutritional supplement, can interfere with VITROS GLU, LAC, TRIG, and URIC Slides assays.