Skip to content
Recall Observatory FDA recall evidence

Device product

PKG, GRASPER FORCEPS, ATRAUMATIC, P/N 0250080323. Laparoscopic Manual Instruments are intended for cutting, holding, grasping and manipulating of tissue, organs, organ areas and surgical auxiliaries such as suturing material. Some instruments have radio frequency (RF) ports which allow them to be used for electrosurgery. Laparoscopic surgical manual instruments are designed to be used endoscopically through cannulae to perform cutting, grasping, dissecting, retracting, and manipulating functions. This device does not sustain or support life.

Z-0686-2015

November 17, 2014

Class II

Product summary

Firm
Stryker Endoscopy
Event
Event 69838
Status
Terminated
Classification
Class II
Quantity
180,573 units total
Official record key
device-enforcement:Z-0686-2015

Official wording

Reason: The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Code information: 747135 840073 840364 846994 1141383 1141658 1142912 1143122 1143498 1144242 1144502 1240415 1241513 1246170 1340543 1341580 1342751 1346234 642223

Distribution pattern: Worldwide Distribution - USA (nationwide) and Internationally to Algeria, Argentina, Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, France, Germany, Greece, Guam, Guatemala, India, Italy, Japan, Jordan, Kuwait, Latin America, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Venezuela.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The parameters provided in the Laparoscopic Manual Instruments Instructions for Use (IFU 1000-401-070 Revision G or prior) do not support the unwrapped Gravity cycle and the Ethylene Oxide cycle sterilization methods

Field note

Send feedback