Skip to content
Recall Observatory FDA recall evidence

Device product

MHI-TM2000 Linear Accelerator System (Software Version 3.5.0 and 3.5.1)

Z-1574-2015

April 17, 2015

Class II

Product summary

Firm
MITSUBISHI HEAVY INDUSTRIES, LTD., HIROSHIMA MACHINERY WORK
Event
Event 70638
Status
Terminated
Classification
Class II
Quantity
11 units total
Official record key
device-enforcement:Z-1574-2015

Official wording

Reason: The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.

Code information: Serial Numbers: 201902, 201903, 202903, 202905, 203901, 203902, 203903, 203905, 203906, 203919 & 203924

Distribution pattern: Worldwide Distribution - US including FL, NY, OH, & TX, **Internationally - Belgium, Italy, Korea, France & Germany.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The operator console allows users to deliver therapeutic radiation to patients even though a specific communication error relevant to imaging conditions has occurred. In addition, the treatment record (the delivered radiation record) cannot be saved.

Field note

Send feedback

We'll only use this to respond to your feedback.