Skip to content
Recall Observatory FDA recall evidence

Device product

S4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with a defined torque 2.8 Nm for tightening the clamping screw in the screw head/hook head/occipital plate/transverse connector and other connectors.

Z-2326-2015

July 07, 2015

Class II

Product summary

Firm
Aesculap, Inc.
Event
Event 71709
Status
Terminated
Classification
Class II
Quantity
10
Official record key
device-enforcement:Z-2326-2015

Official wording

Reason: A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

Code information: product: FW103R, all Lots

Distribution pattern: Distributed US (nationwide) and to Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    A FW103R torque wrench that was improperly reassembled after reprocessing caused a screw to break during implantation due to too much torque. The firm did not include instructions for reassembly after reprocessing.

Field note

Send feedback

We'll only use this to respond to your feedback.