Skip to content
Recall Observatory FDA recall evidence

Device product

TotalCare Bed Systems, Model Numbers P1900, PR1900, and U1900 with 3-Level Patient Position Monitoring and Scale Upgrade Kits, Model Numbers P1947A and P1948A).

Z-1816-2017

February 23, 2017

Class II

Product summary

Firm
Hill-Rom, Inc.
Event
Event 76506
Status
Terminated
Classification
Class II
Quantity
3037
Official record key
device-enforcement:Z-1816-2017

Official wording

Reason: The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.

Code information: P1900 SN: K328AM6065 to Q245AM0691 PR1900 SN: L082AL7316 to R343AL4516 U1900 SN: L257AM0541 to O311AM7169 Scale Upgrade Kit P1957A produced after March 13, 2013 (not serialized).

Distribution pattern: US Nationwide. AUSTRALIA BANGLADESH BRAZIL CANADA CHINA GERMANY INDIA ITALY KUWAIT PAKISTAN SAUDI ARABIA THAILAND UNITED ARAB EMIRATES CYPRUS GREECE NETHERLANDS

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    The bed exit alarm may not function properly if the alarm has been armed for a period longer than seventy-two (72) hours without the alarm being triggered by patient movement or removal of patient weight within that time period.

Field note

Send feedback

We'll only use this to respond to your feedback.