Skip to content
Recall Observatory FDA recall evidence

Device product

Blanketrol III, Model Number 233, Cincinnati Sub-Zero Products, Inc., 12011 Mosteller Road, Cincinnati, OH 45241 Hyper-Hypothermia device used to lower or to raise a patients temperature and/or maintain a desired patient temperature through conductive heat transfer.

Z-2209-2013

September 03, 2013

Class II

Product summary

Firm
Cincinnati Sub-Zero Products Inc
Event
Event 66155
Status
Terminated
Classification
Class II
Quantity
115V - 396 units; 230V - 385 units
Official record key
device-enforcement:Z-2209-2013

Official wording

Reason: During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.

Code information: Part number: 86107/06107/6107 (115V), Serial #s: 112-3-02021, 112-3-02102 through 124-3-02783, 074-3-00388, 081-3-00489, 081-3-00493, 082-3-00511, 091-3-00935, 094-3-01261, 094-3-01265, 094-3-01272, 094-3-01304, 094-3-01308, 101-3-01384, 093-3-01045, 093-3-01046, 093-3-01069, 084-3-00775, 073-3-00315, 091-3-00918, 091-3-00919, 073-3-00279, 094-3-01262, 071-3-00183, 092-3-00983, 094-3-01255, 064-3-00114, 072-3-00219, 072-3-00222, 072-3-00230, 072-3-00258, 081-3-00499, 081-3-00500, 064-3-00116, 064-3-00117, 092-3-01021, 073-3-00305, 073-3-00314, 082-3-22572, 082-3-22573, 101-3-01380, 073-3-00317, 074-3-00341, 091-3-00932, 094-3-01310, 101-3-01374, 082-3-00559, 092-3-00983, 094-3-01255, 094-3-01305, 062-3-00029, 093-3-01048, 094-3-01240, 102-3-01504 & 094-3-01243. Part number: 86207/6207 (230V) Serial #s: 112-3-01949, 114-3-02171 through 124-3-02764, 072-3-00247, 073-3-00266, 073-3-00282, 082-3-00564, 071-3-00132, 071-3-00152, 073-3-00309, 082-3-00533, 082-3-00534, 082-3-00565, 082-3-00568, 084-3-00768, 084-3-00770, 084-3-00777, 084-3-00778, 084-3-00779, 084-3-00780, 084-3-00781, 084-3-00787, 084-3-00788, 091-3-00836, 091-3-00877, 091-3-00878, 091-3-00902, 092-3-01007, 092-3-01008, 092-3-01011, 102-3-01423, 102-3-01426, 103-3-01592, 103-3-01593, 103-3-01594, 103-3-01595, 103-3-01601, 093-3-01072, 093-3-01073, 093-3-01076, 093-3-01078, 093-3-01084, 093-3-01085, 093-3-01086, 093-3-01087, 093-3-01088, 093-3-01104, 093-3-01105, 093-3-01106, 093-3-01107, 093-3-01108, 103-3-01573, 103-3-01574, 103-3-01575, 071-3-00178, 071-3-00180, 074-3-00393, 074-3-00394, 074-3-00395, 074-3-00396, 074-3-00404, 074-3-00405, 091-3-00904, 091-3-00905, 092-3-00939, 092-3-00948, 092-3-00949, 092-3-00951, 092-3-00962, 092-3-00963, 094-3-01178,094-3-01179, 094-3-01180, 094-3-01181, 094-3-01183, 084-3-00789, 084-3-00790, 084-3-00791, 084-3-00812, 084-3-00813, 084-3-00814, 084-3-00815 & 084-3-00816.

Distribution pattern: Worldwide Distribution - USA including AK, AR, AZ, CA, CO, DC, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MT, , NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI & WV. Internationally to ARGENTINA, AUSTRALIA, AZERBAIJAN, BELGIUM, CANADA, CHILE, CHINA, COSTA RICA, CROATIA, GERMANY, HONDURAS, HONG KONG, INDIA, INDONESIA, ISRAEL, KOREA, KUWAIT, MEXICO, NEW ZEALAND, PERU, SAUDI ARABIA, SINGAPORE, SPAIN, TAHITI, THAILAND, THE NETHERLANDS, THE PHILIPPINES, UNITED ARAB EMERITES & UNITED KINGDOM.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    During a label review it was discovered that both the Blanketrol III Operations (57201-N) & Operations & Technical (56201-Q) manuals had verbiage added back into the manual during ECN M111-4260 (released 12/29/2011) which was previously removed for clarification of the intended use and subsequent 501(K) approval.

Field note

Send feedback

We'll only use this to respond to your feedback.