Device product

StatLock CV Plus w/Pigtail, Catheter Stabilization Device Sterile EO Single Use Only, C.R. Bard, Inc. The StatLock¿ device provides stabilization for compatible medical tubes and catheters.

Z-1618-2013

June 10, 2013

Class II

Product summary

Firm
Bard Access Systems
Event
Event 65202
Status
Terminated
Classification
Class II
Quantity
3,875 units.
Official record key
device-enforcement:Z-1618-2013

Official wording

Reason: BARD Access Systems is recalling lot JUWJF283 of StatLock CV Plus Stabilization Butterfly Devices due to incorrect pad assembly used during manufacture. The absence of the hydrocolloid strip on the bottom may pose an incremental risk including localized skin irritation. Missing directional arrow may lead to mis-orientation of the StatLock stabilization which may impact the securement of the cathet

Code information: Lot JUWJF283

Distribution pattern: National and Global distribution (Europe, European Economic Area, Argentina, China, and Brazil).

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    BARD Access Systems is recalling lot JUWJF283 of StatLock CV Plus Stabilization Butterfly Devices due to incorrect pad assembly used during manufacture. The absence of the hydrocolloid strip on the bottom may pose an incremental risk including localized skin irritation. Missing directional arrow may lead to mis-orientation of the StatLock stabilization which may impact the securement of the cathet