Skip to content
Recall Observatory FDA recall evidence

Device product

Smiths Medical, BCI¿ Remote Alarm Cable (Normally Open & Normally Closed), Reorder Numbers: WW3398NO & WW3398NC. For use with respiration carbon dioxide measurement equipment. Product Usage: The BCI¿ Remote Alarm Cable is designed for use with the 9004 Capnocheck¿ Capnography (Plus or Sleep) system to transfer alarm signals from the Capnocheck¿ Capnography system to a remote alarm system. The BCI¿ Remote Alarm Cable and the 9004 Capnocheck¿ Capnography systems are intended for use in all critical environments including ventilatory applications, patient transport, and anesthesia.

Z-1014-2013

February 06, 2013

Class I

Product summary

Firm
Smiths Medical ASD, Inc.
Event
Event 64543
Status
Terminated
Classification
Class I
Quantity
39
Official record key
device-enforcement:Z-1014-2013

Official wording

Reason: In rare instances, the BCI¿ Remote Alarm Cables (BCI¿ Cable) are not transferring alarms when used with some remote nurse alarm systems.

Code information: Lot Numbers: WW3398NO0820, WW3398NO0847, WW3398NO0948, WW3398NC0822, WW3398NC0847, WW3398NC0947.

Distribution pattern: USA Nationwide Distribution - including the states of: CO, IL, IN, MA, SC and WI.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    In rare instances, the BCI¿ Remote Alarm Cables (BCI¿ Cable) are not transferring alarms when used with some remote nurse alarm systems.