The Zimmer Air Dermatome, REF 00- 8801- 001- 00 , Rx, non-sterile grafting instruments intended to provide variable graft thickness and width capabilities

Z-1483-2013

May 07, 2013

Class II

Product summary

Firm
Zimmer Surgical Inc
Event
Event 65258
Status
Terminated
Classification
Class II
Quantity
84 units
Official record key
device-enforcement:Z-1483-2013

Official wording

Reason: This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.

Code information: Serial Number 100211, 100338, 100515, 100560, 100647, 100755, 101104, 101345, 101562, 101769, 101890, 101983, 102560, 103407, 103509, 103816, 104529, 104863, 105049, 105072, 105604, 106394, 106474, 106624, 107861, 108103, 108203, 108210, 108339, 108911, 109445, 109543, 110034, 110837, 111046, 111047, 111048, 111049, 111050, 111051, 111052, 111053, 111054, 111055, 111056, 111057, 111058, 111059, 111060, 111061, 111062, 111063, 111064, 111065, 111066, 111067, 111068, 111069, 111071, 111072, 111073, 111074, 111075, 111076, 111079, 111081, 111083, 111084, 111085, 111086, 111087, 111089, 111090, 111091, 111092, 111096, 111100, 111101, 111102, 111104, 111105, 111116, 111121, 111122

Distribution pattern: Worldwide distribution: USA Nationwide including Puerto Rico; UAE, TAIWAN, SWEDEN, THAILAND, KOREA, SINGAPORE, HONG KONG CHINA, CHILE, CANADA, AUSTRALIA, JAPAN, ENGLAND, GERMANY, and EGYPT.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    This removal is in response to complaints alleging that the Zimmer Air Dermatomes and Zimmer Air Dermatome II products would either not operate or operate intermittently. Zimmer's investigation determined that the planetary gear teeth were broken. This may cause a delay or postponement of the surgery or result in a suboptimal graft retrieval with potential of requiring an additional donor site.