Skip to content
Recall Observatory FDA recall evidence

Device product

Ashland(TM) Box Film Ball Cube II EBT3, P/N: 027863-002 rev E Product is used to calibrate CyberKnife systems

Z-3211-2018

August 31, 2018

Class II

Product summary

Firm
Accuray Incorporated
Event
Event 80933
Status
Terminated
Classification
Class II
Quantity
259 units
Official record key
device-enforcement:Z-3211-2018

Official wording

Reason: the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System

Code information: All units with the Ashland Brand distributed between August 1, 2017 and July 31, 2018

Distribution pattern: Worldwide Distribution: United States (nationwide), and countries of: United Kingdom, Germany, France, Finland, Netherlands, Portugal, Hong Kong, Poland, Australia, United Arab Emirates, Thailand, and Japan.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    the Ball Cube II phantom calibration film was not manufactured within necessary dimensions and may introduce up to 0.5 mm of positional inaccuracy to the CyberKnife System

Field note

Send feedback

We'll only use this to respond to your feedback.