NeuViz 16 Multi-Sliced CT Scanner System, part number 989605858501. NeuViz Dual series CT Scanner System, part number 989605651321 The NeuViz 16 CT Scanner System can be used as a whole body computed tomography X-ray system featuring a continuously rotating X-ray tube and detector array with multi-slice capability up to 16 slices simultaneously.
Z-0466-2013
Product summary
- Event
- Event 63553
- Status
- Terminated
- Classification
- Class II
- Quantity
- 20 units of NeuViz 16 CT Scanner Systems and 25 units of NeuViz Dual series
- Official record key
device-enforcement:Z-0466-2013
Official wording
Reason: The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.
Code information: NeuViz 16 system Serial numbers: N16E090002, N16E090003, N16E090004, N16E090007, N16E090008, N16E090010, N16E100015, N16E100017, N16E100023, N16E110016, N16E110023, N16E110026, N16E110027, N16E110040, N16E110053, N16E120016, N16E120022, N16E120020, N16E120023, N16E120024. Neuviz Dual Serial numbers: 400562, 400563, 400626, NDH009EI, NDH011EI, NDH012EI, NDH013EI, NDH016EI, NDH017EI, NDH026EI, NDH027EI, NDH028EI, NDH030EI, NDH031EI, NDH034EI, NDHR080001, NDHR080002, NDHR080003, NDHR080004, NDHR080007, NDHR080012, NDHR080021, NDHR090003, NDHR090012, and NDHR090013.
Distribution pattern: Nationwide Distribution including North Carolina, Ohio, Texas, and Puerto Rico.
Derived failure modes
-
Unknown
The four screws that secure the patient table top to the carrier pulled out and caused the table top to detach. This issue may affect NeuViz Dual and NeuViz 16 scanners.