Skip to content
Recall Observatory FDA recall evidence

Device product

ROSA Surgical Device 2.5.8, a computer-controlled electromechanical arm to guide standard neurosurgical instruments.

Z-0160-2018

November 06, 2014

Class II

Product summary

Firm
Zimmer Biomet, Inc.
Event
Event 77626
Status
Terminated
Classification
Class II
Quantity
31 units
Official record key
device-enforcement:Z-0160-2018

Official wording

Reason: Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).

Code information: Model No. 2.5.8., Serial No. RO10009, RO10011, RO10014, RO13023, RO13027, RO14031, RO14033, RO14035, and RO14037.

Distribution pattern: US distribution in AR, GA, OH, MI, and TX.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Potential failure of a force feedback sensor that may result in continued movement of an operating arm (rather than an immediate stop as intended).

Field note

Send feedback

We'll only use this to respond to your feedback.