Skip to content
Recall Observatory FDA recall evidence

Device product

Optilite Freelite Lambda Free Kit Product Code: LK018.OPT, LK018.OPT.A Protein testing kit.

Z-1043-2017

July 05, 2016

Class II

Product summary

Firm
The Binding Site Group, Ltd.
Event
Event 76048
Status
Terminated
Classification
Class II
Quantity
809 kits
Official record key
device-enforcement:Z-1043-2017

Official wording

Reason: Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.

Code information: Lot #405566, 405567

Distribution pattern: U.S. distribution to the following; AR, PA, GA, NC, KY, VA, CA. Foreign distribution to the following; Uk, Germany, France, Italy, Spain, Czech Republic, Portugal, Belgium.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Batch numbers listed are reporting Lambda Free Light Chain results >114mg/L that are on average 33% lower than other Optilite Lambda Freelite batches. Therefore users moving to this batch will see results >114mg/L giving lower results and conversely when switching to the next batch results will be seen to shift upwards.

Field note

Send feedback

We'll only use this to respond to your feedback.