Skip to content
Recall Observatory FDA recall evidence

Device product

Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) is recommended for the assessment of CDX2 protein in normal and neoplastic tissues.

Z-1877-2016

April 05, 2016

Class II

Product summary

Firm
Leica Microsystems, Inc.
Event
Event 73802
Status
Terminated
Classification
Class II
Quantity
90 Units
Official record key
device-enforcement:Z-1877-2016

Official wording

Reason: Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.

Code information: 1ml size Novocastr Novocastr Liquid Mouse Monoclonal Antibody CDX2 (CDX2) with lot numbers 6032734, 6033766, 6037153 and 6039088.

Distribution pattern: Distributed in the following states AR, CA, CT, FL, ID, IL, IN, MT, OH, PA, TN and TX, and in Canada.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Testing of retained units indicated that Lot numbers of CDX2 indicated are not stable up to the expiry date on the product labeling.

Field note