VITROS Immunodiagnostic Products Total T4 Reagent Pack, REF/Product Code 874 4468, Unique Identifier No. 10758750005147, IVD --- Ortho-Clinical Diagnostics, For in vitro diagnostic use only. For the quantitative measurement of total thyroxine (T4) in human serum and plasma (EDTA or heparin) using the VITROS ECi/ECiQ Immunodiagnostic Systems, the VITROS 3600 Immunodiagnostic System and the VITROS 5600 Integrated System to aid in the differential diagnosis of thyroid disease.

Z-2302-2016

June 02, 2016

Class II

Product summary

Firm
ORTHO-CLINICAL DIAGNOSTICS
Event
Event 74366
Status
Terminated
Classification
Class II
Quantity
US: 12,015; Foreign: 55 Sales Units to Canada, 223 Sales Units shipped to Bermuda
Official record key
device-enforcement:Z-2302-2016

Official wording

Reason: Ortho-Clinical Diagnostics confirmed that sample results generated from VITROS Immunodiagnostic Products Total T4 Reagent Packs may show a negative bias prior to the open pack storage limit of <= 8 weeks that is located in the Instructions for Use (IFU) for VITROS Immunodiagnostic Products Total T4 Reagent Pack.

Code information: Lot Numbers:1836, 1840, 1845, 1850, 1860, 1870, 1880, 1890, 1900, 1910, 1920, 1930, 1940, 1950, 1968, 1978, 1988, 1998

Distribution pattern: Worldwide Distribution - US (Nationwide including Puerto Rico) and foreign distribution to Canada and Bermuda

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    Ortho-Clinical Diagnostics confirmed that sample results generated from VITROS Immunodiagnostic Products Total T4 Reagent Packs may show a negative bias prior to the open pack storage limit of <= 8 weeks that is located in the Instructions for Use (IFU) for VITROS Immunodiagnostic Products Total T4 Reagent Pack.