Skip to content
Recall Observatory FDA recall evidence

Device product

ANSPACH - 5MM x 11.3MM Fluted Barrel; REF M-5BA-A-G1 Cutting shaping bone including spine and cranium.

Z-2296-2015

April 13, 2015

Class II

Product summary

Firm
The Anspach Effort, Inc.
Event
Event 71212
Status
Terminated
Classification
Class II
Quantity
103 tools.
Official record key
device-enforcement:Z-2296-2015

Official wording

Reason: G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

Code information: REF M-5BA-A-G1 ; ALL lot numbers

Distribution pattern: Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.

Field note

Send feedback

We'll only use this to respond to your feedback.