Skip to content
Recall Observatory FDA recall evidence

Device product

DS/US Proton Feature with Pinnacle3 Software Version 10.0.0 and 14.0.0 Radiation Therapy Planning System Model #459800200621

Z-0745-2016

January 08, 2016

Class II

Product summary

Firm
Philips Medical Systems
Event
Event 73023
Status
Terminated
Classification
Class II
Quantity
3
Official record key
device-enforcement:Z-0745-2016

Official wording

Reason: In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.

Code information: Software Version 10.0.0 and 14.0.0

Distribution pattern: US: Nationwide Distribution in the states of FL, OH, and MO.

Derived failure modes

  • Unknown

    reason.no_named_rule · v1.0.0

    In DS/US proton planning, there is a correction factor used internally to the dose engine that is calculating improperly in some cases. This can result in an incorrect dose calculation.

Field note

Send feedback

We'll only use this to respond to your feedback.