Device product
The Brilliance 64 and Ingenuity CT are Computed Tomography X-Ray Systems intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. These devices may include signal analysis and display equipment, patient, and equipment supports, components and accessories.
Z-2664-2014
Product summary
- Event
- Event 69045
- Status
- Terminated
- Classification
- Class II
- Quantity
- 6
- Official record key
device-enforcement:Z-2664-2014
Official wording
Reason: During a Field Test (a customer external evaluation period), the customer detected a problem in the CIRS 4.0 Beta 2 software. When using the CIRS 4.0 Beta 2 software, during the recon during ready stage of reconstruction, images may be overlapped with, or superimposed on other images. Philips has disabled this feature at all sites evaluating 4.0 Beta 2 software. Philips will issue a software up
Code information: 91003, 30018, 95504, 300004, 300030, 300010, 95130, 95157
Distribution pattern: Worldwide Distribution: US distribution in OH and countries of: Australia, China, Belgium, Israel, and Sweden.
Derived failure modes
-
Device software or design
software. Philips will issue